Research Article

Development of the RP-HPLC Method for Simultaneous Determination and Quantification of Artemether and Lumefantrine in Fixed-Dose Combination Pharmaceutical Dosage Forms

Table 5

Accuracy study.

APIAccuracy level (%)Amount added (μg/mL)Amount recovered (μg/mL)% recoveryMean % recovery% RSD

Artemether8079457955100.1399.790.01
8123809899.69
8057802299.57
100100391003299.9399.930.00
100411003999.98
100431003299.89
1201195012054100.87100.160.01
120731204599.77
120041198599.84

Lumefantrine809709969599.8699.500.51
9692958798.92
9714968899.73
100121431206299.3399.330.22
121401209899.65
121331210399.75
120145561440198.9499.040.09
145531442399.11
145561442199.07

Acceptance criteria: RSD ≤ 2%.