Development of Electron Paramagnetic Resonance Magnet System for
In Vivo
Tooth Dosimetry
Table 3
Magnetic flux density with pole gap extension. The magnetic field and its homogeneity are estimated for 2 cm DSV.
Pole gap (cm)
Magnetic flux density (mT)
Homogeneity (ppm)
18.0
43.5
2,183
18.2
43.0
2,162
18.4
42.6
2,129
19.0
41.2
2,577