Research Article

Guidewire Ablation within the Coronary Venous System for Epicardial or Intramural Ventricular Arrhythmia: A Preclinical Study of Biophysical Characterization

Table 6

Ablation parameters of guidewire ablation in the animal model.

Power setting15 wNo. of guidewire ablation9
Ablation duration60 secondsNo. of steam pop1
Exposed length of the wire tip20 mmTime to steam pop35 seconds
Saline infusion speed2 ml/minLesion maximum diameter3.2 ± 0.3 mm
Baseline impedance170 ± 13.5 ΩLesion minimum diameter2.8 ± 0.5 mm
Impedance drop11.8 ± 3.7 ΩSurvival time10 d