Research Article

Feasibility of Leveraging Consumer Wearable Devices with Data Platform Integration for Patient Vital Monitoring in Low-Resource Settings

Table 2

Results from the Bland-Altman plot.

DevicesMean differenceLimits of agreementOutliers

Fitbit heart rate vs. patient monitor0.77-3.78 to 5.336
Fitbit heart rate vs. manual monitor-1.56-7.23 to 4.113
Xiaomi heart rate vs. patient monitor-0.86-10.56 to 8.8215
Xiaomi heart rate vs. manual monitor-2.12-15.07 to 10.831
Xiaomi SpO2 vs. patient monitor-0.33-3.47 to 2.814