Research Article

Feasibility of Leveraging Consumer Wearable Devices with Data Platform Integration for Patient Vital Monitoring in Low-Resource Settings

Table 3

Results from mean absolute percentage error (MAPE) calculations.

DevicePhilips MX450 patient monitorManual monitoring device

Fitbit heart rate day 12.84%
Fitbit heart rate day 23.40%
Overall Fitbit heart rate3.13%3.63%
Xiaomi heart rate day 16.15%
Xiaomi heart rate day 25.39%
Overall Xiaomi heart rate5.76%6.79%
Xiaomi SpO2 day 11.92%
Xiaomi SpO2 day 20.99%
Overall Xiaomi SpO21.21%