Research Article
Feasibility of Leveraging Consumer Wearable Devices with Data Platform Integration for Patient Vital Monitoring in Low-Resource Settings
Table 3
Results from mean absolute percentage error (MAPE) calculations.
| Device | Philips MX450 patient monitor | Manual monitoring device |
| Fitbit heart rate day 1 | 2.84% | | Fitbit heart rate day 2 | 3.40% | | Overall Fitbit heart rate | 3.13% | 3.63% | Xiaomi heart rate day 1 | 6.15% | | Xiaomi heart rate day 2 | 5.39% | | Overall Xiaomi heart rate | 5.76% | 6.79% | Xiaomi SpO2 day 1 | 1.92% | | Xiaomi SpO2 day 2 | 0.99% | | Overall Xiaomi SpO2 | 1.21% | |
|
|