Research Article
Determination of the Influence of Strut Connector on Stent Deployment
Table 3
Expansion characteristics of Osum and Envision stents from experiment.
| Stent | Diameter of stent (mm) | Stent length before expansion (mm) | Stent length after expansion (mm) | Radial recoil (%) | Fore-shortening (%) | Dog-boning (%) | Before deflation | After deflation | At distal end |
| Osum | 2.963 | 2.903 | 2.961 | 24 | 23.416 | 2.02 | 2.43 | 1.96 | Envision | 3.469 | 3.239 | 3.459 | 28 | 26.229 | 6.63 | 7.62 | 6.32 |
|
|