Research Article
Determination of the Influence of Strut Connector on Stent Deployment
Table 4
Expansion characteristics of stent models with different strut connectors from simulation.
| Stent model | Maximum stress after deflation (MPa) | Maximum radial displacement (mm) | Radial displacement at the middle (mm) | Axial displacement due to expansion (mm) | Radial recoil (%) | Foreshortening (%) | Dog-boning (%) | Before deflation | After deflation |
| Model A | 461 | 0.731 | 0.645 | 0.617 | -0.049 | 5.49 | 1.56 | 1.89 | Model B | 484 | 0.678 | 0.569 | 0.539 | -0.057 | 7.20 | 1.81 | 2.13 | Model C | 484 | 0.685 | 0.581 | 0.558 | -0.044 | 6.84 | 1.40 | 1.63 |
|
|