Quantifying Region-Specific Elastic Properties of Distal Femoral Articular Cartilage: A Shear-Wave Elastography Study
Table 2
Inter- and intrarater reliability of IN, MFC, and LFC stiffness.
| | | Rater DWX in test 1 | Rater DWX in test 2 | Rater LM | ICCa (95% CI) | ICCb (95% CI) |
| IN (L) | (kPa) | | | | 0.91 (0.78-0.96) | 0.96 (0.89-0.98) | MDC (kPa) | 4.29 | 4.44 | 4.60 | SEM (kPa) | 1.55 | 1.60 | 1.66 | IN (R) | (kPa) | | | | 0.98 (0.94-0.99) | 0.97 (0.93-0.99) | MDC (kPa) | 4.67 | 5.15 | 4.78 | SEM (kPa) | 1.69 | 1.86 | 1.73 | MFC (L) | (kPa) | | | | 0.90 (0.77-0.96) | 0.98 (0.95-0.99) | MDC (kPa) | 3.52 | 3.41 | 3.44 | SEM (kPa) | 1.27 | 1.23 | 1.24 | MFC (R) | (kPa) | | | | 0.90 (0.77-0.96) | 0.92 (0.81-0.97) | MDC (kPa) | 4.55 | 4.34 | 4.58 | SEM (kPa) | 1.64 | 1.57 | 1.65 | LFC (L) | (kPa) | | | | 0.87 (0.70-0.96) | 0.86 (0.68-0.94) | MDC (kPa) | 5.47 | 6.19 | 6.14 | SEM (kPa) | 1.97 | 2.23 | 2.22 | LFC (R) | D (kPa) | | | | 0.83 (0.63-0.93) | 0.94 (0.87-0.98) | MDC (kPa) | 4.72 | 5.65 | 6.03 | SEM (kPa) | 1.70 | 2.04 | 2.18 |
|
|
L: left sides; R: right sides; IN: intercondylar notch; MFC: medial femoral condyle; LFC: lateral femoral condyle; SD (kPa): standard deviation of kPa; ICC: intraclass correlation coefficient; 95% CI: 95% confidence interval; MDC (kPa): minimal detectable change of kPa; SEM (kPa): standard error of measurement of kPa; kPa: kilo Pascal. aIntrarater reliability. bInterrater reliability.
|