Research Article
A New Cardiovascular Mock Loop Driven by Novel Active Capacitance in Normal and Abnormal Conditions
Table 2
The numerical and hMCL results.
| @60 bpm | PV | AS | SVR | The proposed | Numerical | hMCL | Numerical | hMCL | Numerical | hMCL |
| (mm Hg) | 131.22 | 133.78 | 153.52 | 156.08 | 156.14 | 157.13 | (mm Hg) | 10.23 | 7.61 | 13.87 | 10.42 | 12.58 | 14.27 | (mm Hg) | 129.11 | 130.27 | 141.19 | 139.37 | 153.18 | 151.35 | (mm Hg) | 74.64 | 72.48 | 78.33 | 75.17 | 89.77 | 87.41 | (ml) | 118.56 | 116.23 | 97.44 | 99.76 | 91.06 | 92.32 | (ml) | 50.04 | 46.41 | 54.24 | 51.14 | 56.61 | 56.13 | (ml) | 64.52 | 67.82 | 38.19 | 41.66 | 31.39 | 34.65 | (ml) | 4.64 | 4.86 | 2.74 | 2.99 | 2.29 | 2.49 | (dyn s/cm5) | 34 | 30.14 | 144 | 140.53 | 144 | 140.53 | (dyn s/cm5) | 1500 | 1703.38 | 1500 | 1703.38 | 2,500 | 2919.1 |
|
|