[Retracted] Detection of Particle Concentration and Particle Size Based on Aerodynamic Particle Size Spectrometer

Table 8

Comparison of concentration measurement results between the equipment and ES632.

TimeConcentration measurement value (mg/m3)
This equipmentES632Relative deviation (%)

21:50:243.6173.7854613.00
21:50:543.4483.6685955.97
21:51:243.5413.5952461.0
21:51:543.4503.6898454.59
22:20:242.1502.97852813.12
22:20:542.0492.5974489.58
22:21:242.0982.9864288.99
22:21:542.0652.59872420.25
23:50:240.2960.189422-0.05
23:50:540.2870.35972814.98
23:51:240.2590.2487225.78
23:51:540.1570.249782-4.59
0:20:240.1590.1543585.89
0:20:540.0590.15478110.31
0:21:240.0140.16724515.88
0:21:540.0240.1597139.43