Research Article

Electrochemical Performance of Ti-Based Commercial Biomaterials

Table 1

Electrochemical parameters from the polarization curves of the different biomaterials evaluated.

BiomaterialsChemical compositionEcorr (mV)Ba (mv/dec)Bc (mV/dec)Icorr (mA/cm2)

Ti CPTi−316.03432313.6 10−4
MexicoTi−230.53882866.6 10−5
SwitzerlandTi−354.22342041.1 10−4
USATi6Al4V−447.42611819.3 10−5
GermanyTi6Al4V−472.52331718.3 10−5
ChinaTi6Al4V−409.92221539.0 10−5
IndiaTi6Al4V−144.03553454.1 10−4
BrazilTi6Al4V−246.41852568.8 10−5