Abstract

Elements-added diamond-like carbon films for biomedical applications were investigated. The aim of this work was to study the effects of the elemental contents (silicon and silicon-nitrogen) in a DLC film on its properties for biomedical applications. Pure DLC, Si-DLC, and Si-N-DLC films were prepared from C2H2, C2H2 : TMS, and C2H2 : TMS : N2 gaseous mixtures, deposited on an AISI 316L substrate using the plasma-based ion implantation (PBII) technique. The structure of films was analyzed using Raman spectroscopy. The chemical composition of films was measured using energy dispersive X-ray spectroscopy (EDS). The average surface roughness of films was measured by using a surface roughness tester. The hardness and elastic modulus of films were measured by using a nanoindentation hardness tester. The friction coefficient of films was determined using a ball-on-disk tribometer. The surface contact angle was measured by a contact angle measurement. The corrosion performance of each specimen was measured using potentiodynamic polarization. The biocompatibility property of films was conducted using the MTT assay cytotoxicity test. The results indicate that the Si-N-DLC film shows the best hardness and friction coefficient (34.05 GPa and 0.13, respectively) with a nitrogen content of 0.5 at.%N, while the Si-DLC film with silicon content of 14.2 at.%Si reports the best contact angle and corrosion potential (92.47  and 0.398 V, respectively). The Si-N-DLC film shows the highest cell viability percentage of 81.96%, which is lower than the uncoated AISI 316L; this is a considerable improvement. All specimens do not demonstrate any cytotoxicity with approximate viabilities between 74% and 107%, indicating good biocompatibilities.

1. Introduction

Implantable biomedical devices are one of the most popular methods of medical field to treat human illnesses. There are many kinds of biomedical devices, such as vascular stents, artificial joints, artificial knees, and bone plates [1]. Most biomedical devices are made from stainless steel and titanium alloys because of the favorable mechanical and biocompatibility properties of these materials [25].

However, problems arising from material deterioration are often detected after long-term use. Corrosion is unavoidable due to various ions in the body reacting electrochemically with the surface of these metallic materials [6]. A metal device may release metal into the body, causing allergic reactions as thrombogenicity of the blood in complicated and aggressive physiological environments. Thus, ideal biomedical devices provide biocompatibility that prevents metallic ion release [6]. Surface coatings can improve both the mechanical properties and biocompatibility of biomedical devices that are in direct contact with blood and tissue.

Coating film technology has been studied for biomedical applications and includes diamond-like carbon (DLC) film. DLC films have many excellent properties including high hardness, low friction coefficient, good corrosion resistance, and good biocompatibility properties [710]. Additionally, DLC films can be doped with certain elements, such as hydrogen (H-DLC), fluorine (F-DLC), and sulfur (S-DLC), to improve performance. Previous studies have shown that DLC films doped with silicon (Si-DLC) are significantly improved corrosion properties due to the formation of a passive film on their surfaces [1116]. Moreover, silicon-doped DLC films deposited by the sputtering method with the Si concentration varied from 4 to 16 at.% can reduce platelet adhesion on the material surface by modifying the hydrophobicity of the material [17]. Additionally, the corrosion resistance of a pure DLC film can be increased via silicon and nitrogen doping by increasing the number of sp3 sites in the film [18].

Plasma-based ion implantation (PBII) has been developed to improve DLC film properties and fabricate three-dimensional (3D) materials with complex shapes. In this technique, low working temperature avoids film quality degradation, such as loose and rough surface structure, and avoids DLC graphitization caused by normal CVD technique which is performed at higher working temperature [19, 20].

Currently, there is no report on the deposition of Si- and Si-N-added DLC films on the AISI 316L substrates by the PBII method aimed at comparing the mechanical, tribological, and corrosion performance, especially to compare the cell viability percentage.

In this paper, plasma-based ion implantation (PBII) was utilized to prepare different elements including silicon- and silicon-nitrogen-added DLC films (henceforth denoted as Si-DLC and Si-N-DLC films). The aim of the study was to compare and study the effects of element contents on the deposition and properties of the films for biomedical applications.

2. Materials and Methods

A schematic of the PBII apparatus used in this study is shown in Figure 1 [18]. The inner dimensions of the vacuum chamber are 600 × 630 × 200 mm3, and the chamber produced a residual pressure of approximately 1 × 10−4 Pa. Plasma was generated from a radio frequency (RF, 13.6 MHz) glow discharge and a negative high-voltage pulse power supply connected to the sample holder.

Medical-grade AISI-type 316L stainless steel was used for the substrate material. Prior to plasma coating, specimens were polished down to 2000 grits using standard abrasive paper and the mirror was polished with 1 μm diamond paste and then cleaned in an ultrasonic cleaning bath. Substrates were sputter-cleaned with Ar+ for 20 min to remove surface contaminants and surface oxides using a −10 kV pulse bias voltage. The DLC film interlayer was then deposited with CH4 for 60 min to improve adhesion between the film and the substrate using a bias voltage of −20 kV. The RF power was set at 300 W, and the pressures of the sputter-cleaning and interlayer deposition were both 1 Pa. A pulse width of 5 μs, a pulse delay of 25 μs, and a pulse frequency of 1 kHz were used during the sputter-cleaning and interlayer deposition processes. Pure DLC, Si-DLC, and Si-N-DLC films were deposited using the PBII technique. The pure DLC film was used with a C2H2 precursor gas. The Si-DLC film was prepared from gaseous mixtures of C2H2 and TMS at three different flow rate ratios (1 : 2, 1 : 4, and 1 : 6). The Si-N-DLC film was deposited from gaseous mixtures of C2H2 : TMS : N2 at three different flow rate ratios (14 : 1 : 2, 14 : 1 : 4, and 14 : 1 : 6). The bias voltage and deposition pressure of all of the films were set at −5 kV and between 2 and 2.5 Pa, respectively. The RF power was set at 300 W, and the pulse frequency was set to 1 kHz at a pulse width of 5 μs and a pulse delay of 25 μs. The thickness of each film was approximately 500 nm. The deposition conditions for the DLC, Si-DLC, and Si-N-DLC films are shown in Table 1.

The structures of the films were analyzed using Raman spectroscopy (JASCO NRS-1000 DT; beam diameter = 4 μm and wavelength = 532 nm). The chemical composition of films was measured using energy dispersive X-ray spectroscopy (EDS). The average surface roughness (Ra) at the top surface of the films was measured using a surface roughness tester (MAHR; MARSURF PS 10; length of travel = 4.8 mm, and cutoff length = 0.8 mm). The hardness and elastic modulus of films were measured using a nanoindentation hardness tester using a diamond ball indenter (Berkovich-type) with an indentation load of 1,000 μN. The friction coefficients of the films were determined using a ball-on-disk tribometer under ambient air conditions. In the friction test, a dry sliding test was performed using a ball indenter made of AISI 440C (SUS 440C, diameter of 6.0 mm) under a normal applied load of 1 N, a rotational radius of 4 mm, a linear speed of 31.4 mm/s, and 3,000 frictional rotations. The surface contact angle was measured by using a contact angle measurement device from Kyowa Interface Science Co., Ltd. using 1 μL of distilled water. The corrosion performance of each specimen was measured using potentiodynamic polarization in a simulated physiological Ringer’s solution (ASTM F2129-17) at a pH of 7.5. A Pt sheet and Ag/AgCl were used as the counter and the reference electrodes, respectively. The potential voltage was varied from −1 V up to +1 V at a scanning rate of 1 mV/s. The cytotoxicity was selected for testing from the most abundant element added to the AISI 316L substrate; these were the pure DLC, Si-DLC 1 : 6, and Si-N-DLC 14 : 1 : 6 based on the ISO 10993-5 guidelines using the Dulbecco’s modified Eagle medium (DMEM) dilution method. All samples were placed in sterile Duran bottles and then sterilized at 121°C for 15 min. L929 fibroblast cells (mouse fibroblasts) at 1 × 105 cells/mL were grown in a 96-well plate at a volume of 100 μL/well and then kept in an incubator (37°C, 5% CO2) for 24 hours. The cell viability was measured using an MTT assay and a microplate reader. This assay was performed in triplicate on cells seeded onto a 96-well plate.

3. Results and Discussion

3.1. Structure of Films by Raman Spectroscopy

Raman spectroscopy was used to investigate the DLC film’s structure due to its nondestructive mechanism. The Raman spectra measured are shown in Figures 24 and indicate that the Raman spectra of pure DLC, Si-DLC, and Si-N-DLC films deposited at different flow rate ratios onto the fabricated AISI 316L substrates. The Raman spectra in the wavelength of 1,000 cm−1–1,800 cm−1 were deconvoluted into Gaussian D and G peaks. The Raman spectrum was measured by curve fitting procedure using two Gaussian distribution peak shapes as show in Figure 2. The peak position and ID/IG ratio of the integrated areas by the D and G band in the DLC films fabricated with various gaseous mixtures and gaseous ratios are presented in Table 2. It has been obtained that the position of G band is related to the bond-angle disorder or sp3 bonding content, while the ID/IG ratio is related to disorder [16, 21]. The films in this experiment show a broad spectrum composed of a D band (1,350 cm−1) and a G band (1,580 cm−1), which are similar to the peaks produced by conventional DLC films.

The results shown in Figures 3 and 4 and Table 2 show that the G peak and ID/IG ratio of pure DLC decrease due to silicon and silicon-nitrogen incorporation. For the Si-DLC film, the G peak tends to decrease from 1467 cm−1 (C : Si/1 : 2) to 1456 cm−1 (C : Si/1 : 6), while the ID/IG ratio decreases from 0.15 to 0.11 as the silicon content increases. This result is likely due to the altered microstructure as more sp3 structures form as the silicon content increases. Additionally, longer destrained bonds vibrate at a lower frequency, leading to the G band shifting to a lower frequency. This outcome can be partially attributed to a reduction in compressive stress when silicon is introduced into the films [12].

For the Si-N-DLC film, the G peak tends to increase slightly from 1521 cm−1 (C : Si : N = 14 : 1: 2) to 1524 cm−1 (C : Si : N = 14 : 1 : 6), while the ID/IG ratio increases from 0.55 to 0.58 due to silicon-nitrogen incorporation. These results were likely caused by unstable sp3 cluster formation and the increasing number or size of the film’s graphitic structure due to nitrogen incorporation. These results thus suggest that the formation of a higher sp2 and lower sp3 clusters in the film is caused by an increase in the silicon-nitrogen content [22].

3.2. Relative Atomic Content of Films

The relative atomic contents measured on the top surface of the films deposited on the stainless-steel substrate are shown in Table 3. The carbon (C), silicon (Si), nitrogen (N), oxygen (O), and iron (Fe) concentrations were measured using the EDS, and values are shown in units of atomic percent (at.%). Concentrations were normalized to a total of 100 at.% and ignored the contribution of hydrogen because the hydrogen content could not be measured by EDS. The results also show that Fe appeared in all films, likely due to the thin film thickness (approximately 500 nm). It is assumed that the penetration depth of the electrons during the test is higher than the film thickness [18].

As shown in Table 3, the DLC film exhibits the highest C content (91.1 at.%), while the Si-DLC film shows that the Si content increased from 13.3 to 14.2 at.% by changing the gas flow rate from 1 : 2 to 1 : 6. The concentration of the Si-N-DLC film shows that the N content increased from 0.2 to 1.0 at.% when the gas flow rate changed from 14 : 1:2 to 14 : 1:6. In addition, a part of the EDS broad spectrum of Si-N-DLC film at the ratio of 14 : 1:6 is shown in Figure 5.

3.3. Surface Hardness and Elastic Modulus of Films

The hardness and elastic modulus of a film are important for biomedical application because contact pressures in some applications, such as artificial knees and joint replacements, may cause excessive material wear. The wear resistance of a material is directly related to its hardness and elastic modulus.

Each film’s hardness and elastic modulus were determined on its top surface using a nanoindentation hardness test and a diamond ball indenter (Berkovich-type) with an indentation load of 1,000 μN. The hardness and elastic modulus of pure DLC, Si-DLC, and Si-N-DLC are shown in Figures 6 and 7 and Table 4.

As shown in Table 4, the hardness and elastic modulus of the films increased from those of pure DLC (25.46 GPa and 184.51 GPa, respectively) as the silicon content increased. For Si-DLC, the hardness and elastic modulus increased from 29.16 GPa (13.3 at.%Si) to 31.15 GPa (13.7 at.%Si) and 209.76 GPa to 216.92 GPa when the TMS flow rate ratio increased. The increased hardness and elastic modulus due to the change in microstructure was observed via the Raman analysis; the formation of sp3 site structures increased the hardness and elastic modulus as the silicon content increased [18, 23]. However, the hardness and elastic modulus of the 1 : 6 Si-DLC film (14.2 at.%Si) decreased to 26.05 and 185.85 GPa, respectively. C2H2 and TMS used during fabrication: when the TMS flow rate increased, the hydrogen content in the film increased. The polymeric sp3 C−H bonds established due to the high hydrogen concentration decreased the film’s hardness [23]. Si-DLC films are likely to be saturated with hydrogen, especially when using a TMS precursor. The same trend verifies the reliability of our results.

The Si-N-DLC film exhibited the high hardness and elastic modulus values 34.05 GPa and 220.97 GPa (0.5 at.%N), respectively, despite its number and size of sp3 clusters being lower than those of the Si-DLC film. The formation of C−N bonds yielded high hardness and elastic modulus values because C−Si bonds are weaker than C−N bonds. However, when the nitrogen was increased to 1.0 at.%N, the hardness and elastic modulus decreased to 26.28 GPa and 185.61 GPa, respectively, which was likely due to the lower bond strength of the N−Si network [24, 25].

3.4. Friction Coefficient of Films

The friction coefficients of pure DLC, Si-DLC, and Si-N-DLC films were determined under ambient air conditions and are shown in Figure 8. The effects of the silicon and silicon-nitrogen contents on the friction coefficients of the Si-DLC and Si-N-DLC films were investigated. The results indicate that all coating films can increase the low friction coefficient of AISI 316L.

The friction coefficient of pure DLC was 0.17. It was found that increasing the silicon content in the film increased the film’s friction coefficient. The Si-DLC film’s friction coefficient increased from 0.22 (13.3 at.%Si) to 0.30 (14.2 at.%Si) when the TMS flow rate ratio increased, which reduced the carbon content in the film. In general, carbon acts as a solid lubricant that reduces friction between ball-film contact areas [26, 27]. Additionally, for the Si-DLC film, the friction coefficient also increased due to the film’s lower surface hardness and elastic modulus. The contact area at the ball-film interface actually increased as the film hardness and elastic modulus decreased, resulting in the film surface having difficulty supporting the high load induced by the friction coefficient [28]. The Si-N-DLC film exhibited the lowest friction coefficient (0.13 at 0.5 at.%N) due to the film’s high hardness and elastic modulus. The film’s low friction coefficient can be attributed to the lubricating effect of the sp2 cluster in the film coatings, as shown in Table 2 [29].

3.5. Contact Angle of Films

The contact angles measured on the top surfaces of the films are shown in Table 5. A volume of 1 μL of distilled water was released onto the top surface of the films under atmospheric conditions at room temperature.

Many studies have shown that surface roughness and surface hydrophilicity strongly increase the blood compatibility of implants: a smooth and hydrophobic surface is beneficial to the blood compatibility [21, 30, 31]. Contact-angle studies play an important role in understanding the surface phenomena of the substrates.

The polished smooth surface of AISI 316L reported a contact angle value of 93.47°, describing the hydrophobic nature of AISI 316L [32, 33]. The contact angle value reduced to 79.33° for the pure DLC film. For the Si-DLC film, the contact angle increased from 88.93° (13.3 at.%Si) to 92.47° (14.2 at.%Si) as silicon content increased. This was likely due to the decreasing surface energy of the films as the silicon content increased and more strongly influenced the hydrophobic property of the film [34]. Moreover, the contact angle was relatively unaffected by surface roughness, as shown in Table 3.

For the Si-N-DLC film, the contact angle was found to be related to the nitrogen and carbon atomic concentration ratio. The contact angle decreased as the nitrogen content increased, as shown in Table 5. The contact angle of the Si-N-DLC film shifted slightly as the gas flow rate changed from 14 : 1 : 2 to 14 : 1 : 6. The Si-N-DLC surface was thus more hydrophilic than the pure DLC and Si-DLC films, which might be due to an increase in the dispersive and polarity components of surface energy from the high number of nucleophilic electrons produced by the formation of N–H and C=N networks [35].

3.6. Corrosion Resistance Property of Films

Biocompatibility is simply defined as the performance of a material to be permitted by the body. A foreign body is created when materials are implanted in a body and is directly related to the corrosion behavior of the material due to the tendency for the alloy to release potential toxic ions from corrosion behavior [36].

In general, samples with greater corrosion potential and lower current density provide better corrosion performance. An improvement in the corrosion resistance of a film is demonstrated by a shift in the polarization curve towards a region of lower current density and greater potential [37]. Additionally, variation in the film microstructure allows different corrosion behaviors to occur [38]. Generally, a more prevalent sp3 structure tends to increase electrochemical corrosion resistance [18, 38].

Figures 9 and 10 show the potentiodynamic curves for uncoated AISI 316 L, pure DLC, Si-DLC, and Si-N-DLC films. Table 5 summarizes the electrochemical parameters measured from the potentiodynamic polarization curves shown in Figures 9 and 10. All films show higher corrosion potential (Ecorr) and lower current density (icorr) than the uncoated AISI 316 L sample. Pure DLC shows a corrosion potential and current density of −0.010 V and 0.949 × 10−6 A, respectively. Corrosion potential increased, and current density decreased due to silicon (Si-DLC) and silicon-nitrogen (Si-N-DLC) incorporation. For the Si-DLC film, corrosion potential (Ecorr) increased from 0.198 V to 0.398 V and current density (icorr) decreased from 0.887 × 10−6 A/cm2 to 0.280 × 10−6 A/cm2 due to silicon incorporation. For the Si-N-DLC film, corrosion potential (Ecorr) increased from −0.001 V to 0.198 V and current density (icorr) decreased from 0.931 × 10−6 A/cm2 to 0.570 × 10−6 A/cm2 due to silicon-nitrogen incorporation. Thus, the best corrosion resistance was observed in the Si-DLC film with a silicon content of 14.2 at.%Si with corrosion potential (Ecorr) and current density (icorr) values of 0.398 V and 0.280 × 10−6 A/cm2, respectively. The greater corrosion resistance of the Si-DLC and Si-N-DLC films than that of the uncoated AISI 316L and pure DLC film shows that the incorporation of silicon and silicon-nitrogen into the pure DLC film improves corrosion resistance. This outcome is likely due to the increased number of sp3 sites in the film’s structure [39] and the creation of a silicon oxide passive film on the film’s surface [11]. Additionally, the sp3 structure created via silicon incorporation can reduce the internal stress that causes the corrosion [12]. Additionally, the results show more corrosion-resistant potential than the previous report mentioned above [14]. The report showed the corrosion potential and current density of Si-DLC up to 1.5 × 10−5 mA/cm2 and −0.7 V with high silicon content of 30 at.% [14].

3.7. Cytotoxicity Test of Films

A cytotoxic test was designed according to ISO 10993-5 standard, and L 929 fibroblast cells were used in this method. The cytotoxic test is typically used to evaluate materials for biomedical applications to confirm their implantability. The cell viability percentage of uncoated AISI 316L, pure DLC, Si-DLC, and Si-N-DLC are shown in Figure 11.

Figures 11 and 12 show the viability and morphology of the L929 fibroblast cells after being exposed to the MTT solution. The Si-N-DLC film exhibited a cell viability of 81.96%, which is lower than the uncoated AISI 316 L (107%) but higher than the pure DLC (78.16%) and the Si-DLC films (74.4%). However, none of the specimens were found to be cytotoxic. Cell viability ranged from 74.4% to 107%, indicating good biocompatibility because viability percentages (live/dead %) were not lower than 50% as per ISO10993-5. Additionally, no visible change in cell morphology was observed; the cells still exhibited a typical fibroblastic morphology, as shown in Figure 12 [40]. These results agree with those reported by Sui et al. [21], Gotzmann et al. [41], Bociaga et al. [42], and Antunes et al. [43].

4. Conclusions

Elements-added diamond-like carbon films were investigated for biomedical applications. Pure DLC, Si-DLC, and Si-N-DLC films were prepared on an AISI 316L substrate at different flow rate ratios and gaseous mixtures using the plasma-based ion implantation (PBII) technique. All films were described in terms of structure, relative atomic content, surface roughness, hardness, elastic modulus, friction coefficient, contact angle, corrosion resistance, and cytotoxicity. Hardness, elastic modulus, friction coefficient, contact angle, and corrosion resistance increased as the silicon content in the Si-DLC film decreased. Increasing the silicon-nitrogen content in the Si-N-DLC film also increased the hardness and corrosion resistance of the film but decreased the friction coefficient and contact angle. Changes in film hardness, friction coefficient, and corrosion resistance were related to changes in sp3 site structures. Chemical bond strength also affected the hardness and elastic modulus of the films. The Si-N-DLC film showed the best hardness and friction coefficient (34.05 GPa and 0.13, respectively) with a nitrogen content of 0.5 at.%N, while the Si-DLC film with silicon content of 14.2 at.%Si reported the best contact angle and corrosion potential (92.47 and 0.398 V, respectively). The Si-N-DLC film showed the highest cell viability percentage of 81.96%, which was lower than the uncoated AISI 316L; this is a considerable improvement. The DLC and Si-DLC films showed cell viability percentages of 78.16% and 74.4%, respectively. All specimens did not demonstrate any cytotoxicity with approximate viabilities between 74% and 107%, indicating good biocompatibilities.

Data Availability

No data were used to support this study.

Disclosure

A part of this manuscript abstract has been presented in ICASS 2017-Oxford Abstracts.

Conflicts of Interest

The authors declare that they have no conflicts of interest.