Nucleus Disc Replacement: Design and Material Selection FEA Analysis

Table 3

The vertebral endplates’ maximum contact pressures with the selected design nucleus replacement made of Bionate® 80 A, 90 A, and 55D.

Vertebral endplate contact pressures with nucleus replacement with Bionate®80 A90 A55D

Maximum contact pressures 1000 N axial compression loadSmax (MPa)0.7190.981.09
Maximum contact pressures 1000 N axial compression load + 300 N anteroposterior shearSmax (MPa)1.1591.6141.935
Maximum contact pressures 1000 N axial compression load + 8 Nm flexion moment + 8 Nm rotation torqueSmax (MPa)1.2621.7012.033