Nucleus Disc Replacement: Design and Material Selection FEA Analysis
Table 3
The vertebral endplates’ maximum contact pressures with the selected design nucleus replacement made of Bionate® 80 A, 90 A, and 55D.
| Vertebral endplate contact pressures with nucleus replacement with Bionate® | 80 A | 90 A | 55D |
| Maximum contact pressures 1000 N axial compression load | Smax (MPa) | 0.719 | 0.98 | 1.09 | Maximum contact pressures 1000 N axial compression load + 300 N anteroposterior shear | Smax (MPa) | 1.159 | 1.614 | 1.935 | Maximum contact pressures 1000 N axial compression load + 8 Nm flexion moment + 8 Nm rotation torque | Smax (MPa) | 1.262 | 1.701 | 2.033 |
|
|