The Impact of Multiple Autoclave Cycles on the Surface Roughness of Thermally Treated Nickel-Titanium Endodontic Files

Table 2

Comparison of SR parameters between the instruments (means ± standard deviation).

Cycle01510

Ra
TN17.00 ± 4.7a9.68 ± 1.80a11.36 ± 2.80a12.92 ± 1.12a
PTG16.52 ± 2.84a8.33 ± 1.66a8.80 ± 1.65a8.23 ± 0.83b
ETP15.99 ± 5.14a8.16 ± 0.88a11.66 ± 1.47a10.19 ± 1.54b
Rq
TN22.51 ± 6.057a13.13 ± 2.28a15.62 ± 3.53a17.84 ± 1.38a
PTG22.46 ± 4.13a11.44 ± 2.25a11.96 ± 2.15a11.51 ± 1.04b
ETP21.04 ± 6.50a11.28 ± 1.05a16.21 ± 2.05a13.81 ± 1.85b
Rz
TN51.84 ± 21.20a39.81 ± 8.54a45.28 ± 9.81a68.26 ± 7.25a
PTG64.85 ± 12.34a35.70 ± 12.00a35.99 ± 6.95a38.60 ± 7.38b
ETP63.88 ± 19.32a43.89 ± 8.85a82.09 ± 22.22b55.69 ± 3.05a

a, b, and c indicate groups with the same letters have no significant differences. 0 cycle is the pre-autoclaving cycle.