Research Article

Protocol for Evaluation of Robotic Technology in Orthopedic Surgery

Table 1

Translational and rotational error values in UKA implant placement .

TibiaFemoral
Translational error (mm)Rotational error (°)Translational error (mm)Rotational error (°)
Lateral medial Anterior posteriorDistal proximalFlexion extentionVarus valgus Axial rotationLateral medialAnterior posteriorDistal proximal Flexion extentionVarus valgus Axial rotation

Mean0.0−0.90.82.1−0.80.40.6−1.5−1.10.4−0.50.9
SD1.50.51.10.61.61.40.91.41.03.70.92.6
Max1.8−0.22.12.22.23.11.90.4−0.15.30.55.3
Min−2.8−1.6−1.7−3.3−1.3−2.4−1.2−3.2−3.5−4.7−1.9−2.1