A Novel Spine Fixation System Made Entirely of Carbon-Fiber-Reinforced PEEK Composite: An In Vitro Mechanical Evaluation
Table 2
Fatigue axial compression bending test results of the CarboClear system.
| Sample | Load (N) | Percentage of ultimate bending load (%) | No. of cycles to failure | Mode of failure |
| 1 | 200 | 55 | 5 × 106 | No failure, test terminated voluntarily | 2 | 210 | 58 | 7.7 × 106 | No failure, test terminated voluntarily | 3 | 230 | 63 | 5 × 106 | No failure, test terminated voluntarily | 4 | 277 | 76 | 5 × 106 | No failure, test terminated voluntarily | 5 | 300 | 83 | 6 × 106 | No failure, test terminated voluntarily | 6 | 343 | 95 | 150 | Slippage of the polyaxial screw head within the screw tulip |
|
|