A Novel Spine Fixation System Made Entirely of Carbon-Fiber-Reinforced PEEK Composite: An In Vitro Mechanical Evaluation
Table 3
Static torsion test results of the CarboClear system.
| Sample | Yield torque (Nm) | Torsional stiffness (Nm/deg) | Angular displacement at 2% offset yield (deg) | Elastic angular displacement (deg) |
| 1 | 12.4 | 0.9 | 16.1 | 14.2 | 2 | 10.4 | 1.1 | 11.1 | 9.2 | 3 | 12.4 | 1.2 | 12.4 | 10.2 | 4 | 15.1 | 0.9 | 18.3 | 16.2 | Mean ± SD | 12.6 ± 1.9 | 1.0 ± 0.1 | 14.4 ± 3.3 | 12.4 ± 3.3 |
|
|