Table 4
Screw axial pull-out test results of CarboClear pedicle screws (6.5 mm diameter, 35 mm length).
Sample
Axial pull-out strength (N)
1
2151
2
1901
3
2041
4
2057
Mean ± SD
2037 ± 103