Clinical Study
Intubation Success through I-Gel® and Intubating Laryngeal Mask Airway® Using Flexible Silicone Tubes: A Randomised Noninferiority Trial
Table 3
Bootstrapped mean (95% CI) for device insertion and intubation parameters.
| | 
 |  | Variable | Group G () BS mean (95% CI)
 | Group I () BS mean (95% CI)
 | value (sig < 0.05) |  | 
 |  | Number of SGA attempts | 1.20 (1.07; 1.36) | 1.04 (1.00; 1.08) | 0.088 |  | Ease of insertion | 1.40 (1.22; 1.60) | 1.24 (1.14; 1.35) | 0.180 |  | Ease of intubation | 1.66 (1.43; 1.89) | 1.31 (1.19; 1.46) | 0.023 |  | Time to SGA placement (min) | 10.27 (9.07; 11.65) | 9.60 (8.80; 10.44) | 0.418 |  | Time to intubation (min) | 13.39 (11.41; 15.39) | 10.11 (8.76; 11.53) | 0.012 |  | Number of intubation attempts | 1.51 (1.29; 1.76) | 1.31 (1.17; 1.48) | 0.191 |  | 
 | 
 | 
| SGA: supraglottic airway device;  indicates . 
 |