Table 2
Classifier parameters for cell and signal selection.
| Parameters | Variability of parameters | Finally optimized parameters |
| Cell selection parameters | | |
| Minimum cell nucleus area | 0 to 32,000 μm2 | 120 μm2 | Maximum cell nucleus area | 0 to 32,000 μm2 | 500 μm2 | Maximum concavity depth | 0 to 1 | 0.35 | Maximum aspect ratio | 1 to 10 | 2.5 |
| Signal parameters | | |
| Min. absolute signal area | 0 to 100 μm2 | 0.1 μm2 | Max. distance of fused signals | 0.1 to 100 μm | 0.8 μm | Min. of max. signal intensity | 0 to 100% | 33% |
|
|