New Imaging Strategies Using a Motion-Resistant Liver Sequence in Uncooperative Patients

Table 1

Parameters for the motion-resistant protocol used for 3.0T MR imaging scanners.

ParameterPrecontrast sequencesPostcontrast sequences
T1-weighted 2D MP-RAGE in-phase/out-of-phaseT2-weighted SS-ETSEBLADET1-weighted 3D-GRE with CAIPIRINHAT1-weighted 3D-GRE with CAIPIRINHA (echo-sharing technique)T1-weighted radial 3D-GRET1-weighted radial 3D-GRE (KWIC technique)

Plane of acquisitionAxialAxial, coronalAxialAxial, coronalAxialAxialAxial
TR (milliseconds)1500120039104.64.73.43.1
TE (milliseconds)2.3/3.41501052.21.51.71.5
TI (milliseconds)1200
Flip angle (°)2012016012121212.5
Echo train length1022312
Parallel imagingGRAPPA 2GRAPPA 2GRAPPA 22D CAIPIRINHA
2 × 2
2D CAIPIRINHA 2 × 2GRAPPA 2GRAPPA 2
BW/pixel (Hz)210/1807103003958708701040
Matrix (frequency × phase)256 × 173320 × 213320 × 320384 × 227384 × 211384 × 384320 × 320
FOV (mm)302 × 304302 × 394380 × 380302 × 227296 × 299340 × 340340 × 340
Number of sections36363680808080
Section thickness (mm)5552.52.52.53.0
Intersectional gap (mm)1.51.51.50000
Number of signal acquisitions111111
Fat suppressionNoneFat sat for axial, none for coronalFat satFat satFat satFat satFat sat
Respiratory controlBIBIFBBHBHFBBH

BH, breath-hold; BI, breathing-independent; BLADE, proprietary name for periodically rotated overlapping parallel lines with enhanced reconstruction (PROPELLER) in MR systems from Siemens Medical Solutions; BW, bandwidth; Hz, Hertz; CAIPIRINHA, controlled aliasing in parallel imaging results in higher acceleration; FB, free breathing; FOV, field of view; GRE, gradient recalled echo; KWIC, k-space weighted image contrast; MP-RAGE, magnetization-prepared rapid-acquisition gradient echo; sat, saturation; SS-ETSE, single-shot echo-train spin-echo; TE, echo time; TI, inversion time; TR, repetition time.