Research Article

Evaluation of Three Automated Nucleic Acid Extraction Systems for Identification of Respiratory Viruses in Clinical Specimens by Multiplex Real-Time PCR

Table 4

Analytical sensitivity of the three automated nucleic acid extraction systems.

Virus strain
dilution factor
ADV-3 (detected/replicates)FluA (detected/replicates)RSV-A (detected/replicates)

QIAcube (QIAGEN)5/53/50/50/50/55/52/51/50/50/55/54/50/50/50/5
EZ1 (QIAGEN)5/55/50/50/50/55/55/52/51/50/55/55/53/51/50/5
NIMBUS (Seegene)5/54/50/50/50/55/54/50/50/50/55/52/51/50/50/5