Evaluation of Three Automated Nucleic Acid Extraction Systems for Identification of Respiratory Viruses in Clinical Specimens by Multiplex Real-Time PCR
Table 4
Analytical sensitivity of the three automated nucleic acid extraction systems.
Virus strain
dilution factor
ADV-3 (detected/replicates)
FluA (detected/replicates)
RSV-A (detected/replicates)
QIAcube (QIAGEN)
5/5
3/5
0/5
0/5
0/5
5/5
2/5
1/5
0/5
0/5
5/5
4/5
0/5
0/5
0/5
EZ1 (QIAGEN)
5/5
5/5
0/5
0/5
0/5
5/5
5/5
2/5
1/5
0/5
5/5
5/5
3/5
1/5
0/5
NIMBUS (Seegene)
5/5
4/5
0/5
0/5
0/5
5/5
4/5
0/5
0/5
0/5
5/5
2/5
1/5
0/5
0/5