Research Article
Balancing Rigidity and Safety of Pedicle Screw Fixation via a Novel Expansion Mechanism in a Severely Osteoporotic Model
Table 1
Mean and standard deviation for all mechanical parameters recorded.
| | Trabecular block | Trabecular block (with PMMA augmentation) | Composite block | | Pullout (N) | Stiffness (N/mm) | Energy to failure (N*mm) | Energy to removal (N*mm) | Pullout (N) | Stiffness (N/mm) | Energy to failure (N*mm) | Energy to removal (N*mm) | Pullout (N) | Stiffness (N/mm) | Energy to failure (N*mm) | Energy to removal (N*mm) |
| Standard | | | | | | | | | | | | | Mean | 339 | 171 | 425 | 1300 | 338 | 161 | 499 | 2570 | 303 | 177 | 318 | 942 | Standard deviation | 9 | 7 | 21 | 21 | 69 | 12 | 208 | 229 | 10 | 25 | 61 | 133 | Expandable | | | | | | | | | | | | | Mean | 321 | 160 | 396 | 1650 | 360 | 164 | 529 | 3200 | 293 | 179 | 336 | 1380 | Standard deviation | 10 | 11 | 33 | 68 | 28 | 6 | 86 | 527 | 31 | 21 | 94 | 199 |
|
|