X-Ray Scatter Correction on Soft Tissue Images for Portable Cone Beam CT
Table 3
CT number value comparison in the QRM-ConeBeam phantom.
| | FBCT | No correction | Scatter correction at different iterations | | | | | | 2 iterations | 4 iterations | 6 iterations | 8 iterations | 10 iterations | | HU | STD | HU | STD | HU | STD | HU | STD | HU | STD | HU | STD | HU | STD |
| Bone | 752 | 122 | 141 | 82 | 436 | 105 | 648 | 128 | 713 | 139 | 736 | 145 | 742 | 148 | Air | −1000 | 63 | −760 | 57 | −809 | 76 | −905 | 77 | −923 | 70 | −940 | 67 | −943 | 66 | Soft tissue | 0 | 112 | −197 | 63 | −60 | 85 | −2 | 108 | 6 | 117 | 7 | 120 | 7 | 121 |
|
|