Research Article
X-Ray Scatter Correction on Soft Tissue Images for Portable Cone Beam CT
Table 6
HU values of different inserts in Section A of the QRM-ConeBeam phantom.
| | FBCT | CBCT without correction | CBCT with scatter correction | | Mean (HU) | Mean (HU) | Different value (HU) | Mean (HU) | Different value (HU) |
| A1 | −165 | −269 | 104 | −187 | 22 | A2 | −91 | −223 | 132 | −112 | 21 | A3 | −69 | −210 | 141 | −90 | 21 | A4 | −45 | −195 | 150 | −68 | 23 |
|
|