Accuracy of Cardiac Output by Nine Different Pulse Contour Algorithms in Cardiac Surgery Patients: A Comparison with Transpulmonary Thermodilution

Table 4

Correlation and Bland-Altman analysis showing bias, standard deviation from bias, 95% limits of agreement, and percentage error during passive leg raising for nine different pulse contour algorithms compared with transpulmonary thermodilution.

PLR
X1X2X3X4X5X6X7X8X9

Mean (L/min)2.702.722.772.872.832.902.792.832.83
Bias (L/min)−0.54−0.52−0.47−0.37−0.40−0.34−0.45−0.41−0.40
SD of bias (L/min)1.081.131.150.891.171.201.131.151.11
Correlation ()0.180.150.170.360.170.180.170.170.20
95% limits of agreement (L/min)−2.66–+1.59−2.75–+1.72−2.72–+1.77−2.12–+1.38−2.69–+1.89−2.70–+2.02−2.67–+1.76−2.66–+1.84−2.58–+1.75
Percentage error (%)727577587778747573

PLR, passive leg raising; , cardiac output by nine different pulse contour algorithms.