Accuracy of Cardiac Output by Nine Different Pulse Contour Algorithms in Cardiac Surgery Patients: A Comparison with Transpulmonary Thermodilution

Table 5

Polar analysis showing concordance rates with exclusion zone of 15% before (T1) and after (T2) cardiopulmonary bypass for nine different pulse contour algorithms compared with transpulmonary thermodilution.

Before CPB (T1) After CPB (T2)
X1X2X3X4X5X6X7X8X9X1X2X3X4X5X6X7X8X9

Mean (L/min)−0.22−0.29−0.31−0.31−0.32−0.32−0.26−0.27−0.270.741.030.971.120.330.160.170.150.17
Concordance rates (%), 10% limits of agreement596676717175566760614268574261685863
Concordance rates (%), 20% limits of agreement888788949393808680888384798483787689

CPB, cardiopulmonary bypass; , cardiac output by nine different pulse contour algorithms; , changes in cardiac output.
Data pairs within the 10% limits of agreement (10% LOAS: ±0.38 L/min before and ±0.57 L/min after CPB) indicated good trending. Data pairs within the 20% limits of agreement (20% LOAs: ±0.76 L/min before and ±1.14 L/min after CPB) indicated moderate trending ability. Concordance rates >92% were considered sufficient.