The Accuracy of New and Aged Mechanical Torque Devices Employed in Five Dental Implant Systems

Table 2

Accuracies of new and aged wrenches.

Implant
system
Target
value
t95% confidence interval of the differenceMean
difference
p value
LowerUpper

Biohorizons (used)30−11.363−7.4400−5.1527−6.29642p < 0.05
Biohorizons (new)30−17.090−7.2347−5.6755−6.4551p < 0.05

Bredent (used)300.358−2.57093.64930.5391p < 0.05
Bredent (new)30−0.369−2.35131.6386−0.3563p < 0.05

Implant KA (used)25−3.885−2.8184−0.8626−1.8405p < 0.05
Implant KA (new)25−8.605−3.2377−1.985−2.6114p < 0.05

Straumann (used)35−8.754−6.3950−3.9549−5.1749p < 0.05
Straumann (new)35−7.132−6.2778−3.4599−4.8689p < 0.05

Zimmer (used)30−5.906−5.1007−2.4588−3.7798p < 0.05
Zimmer (new)30−2.784−3.7592−0.5582−2.1587p < 0.05