The Accuracy of New and Aged Mechanical Torque Devices Employed in Five Dental Implant Systems
Table 2
Accuracies of new and aged wrenches.
| Implant system | Target value | t | 95% confidence interval of the difference | Mean difference | p value | Lower | Upper |
| Biohorizons (used) | 30 | −11.363 | −7.4400 | −5.1527 | −6.29642 | p < 0.05 | Biohorizons (new) | 30 | −17.090 | −7.2347 | −5.6755 | −6.4551 | p < 0.05 |
| Bredent (used) | 30 | 0.358 | −2.5709 | 3.6493 | 0.5391 | p < 0.05 | Bredent (new) | 30 | −0.369 | −2.3513 | 1.6386 | −0.3563 | p < 0.05 |
| Implant KA (used) | 25 | −3.885 | −2.8184 | −0.8626 | −1.8405 | p < 0.05 | Implant KA (new) | 25 | −8.605 | −3.2377 | −1.985 | −2.6114 | p < 0.05 |
| Straumann (used) | 35 | −8.754 | −6.3950 | −3.9549 | −5.1749 | p < 0.05 | Straumann (new) | 35 | −7.132 | −6.2778 | −3.4599 | −4.8689 | p < 0.05 |
| Zimmer (used) | 30 | −5.906 | −5.1007 | −2.4588 | −3.7798 | p < 0.05 | Zimmer (new) | 30 | −2.784 | −3.7592 | −0.5582 | −2.1587 | p < 0.05 |
|
|