The Accuracy of New and Aged Mechanical Torque Devices Employed in Five Dental Implant Systems
Table 3
Differences between mean values and target values in N⋅cm and as percentages.
| | Biohorizons | Bredent | Implant KA | Straumann | Zimmer |
| Target value (N⋅cm) | 30 | 30 | 25 | 35 | 30 |
| All wrenches | | | | | | Mean value | 23.624 | 30.091 | 22.774 | 29.978 | 27.031 | Difference | −6.376 | 0.091 | −2.226 | −5.022 | −2.969 | Difference% | −21.253 | 0.305 | −8.904 | −14.348 | −9.898 |
| Aged wrenches | | | | | | Mean value | 23.704 | 30.539 | 23.159 | 29.825 | 26.220 | Difference | −6.296 | 0.539 | −1.841 | −5.175 | −3.780 | Difference% | −20.988 | 1.797 | −7.362 | −14.786 | −12.599 |
| New wrenches | | | | | | Mean value | 23.545 | 29.644 | 22.389 | 30.131 | 27.841 | Difference | −6.455 | −0.356 | −2.611 | −4.869 | −2.159 | Difference% | −21.517 | −1.188 | −10.446 | −13.911 | −7.196 |
|
|