Biomechanical Evaluation of a Novel Apatite-Wollastonite Ceramic Cage Design for Lumbar Interbody Fusion: A Finite Element Model Study

Table 4

Maximum stress values of the different cage designs. The stress values were significantly lower in the K2 design ().

 FlexionExtensionBendingRotation
Max.
stress value
Max.
stress point
Max.
stress value
Max.
stress point
Max.
stress value
Max.
stress point
Max.
stress value
Max.
stress point

K1112.1Anterior61.6Posterior100.8Anterolateral72.3Anterior
K232.6Anterior15.3Posterior33.7Anterolateral24.3Anterior
K3114.6Anterior68.4Posterior112.1Anterolateral80.6Anterior
K446.3Anterior20.2Anterior48.6Anterolateral36.8Anterior