Research Article

Reliability and Accuracy of Peri-Interventional Stenosis Grading in Peripheral Artery Disease Using Color-Coded Quantitative Fluoroscopy: A Phantom Study Comparing a Clinical and Scientific Postprocessing Software

Table 1

Results of the ANOVA analysis for ROIs II to IV as analyzed with ImageJ. Percentage changes of contrast intensity curves (area under the curve) using four different integration times (1, 3, 5, and 7 seconds) are compared to a reference ROI 5 cm proximal to the stenosis (ROI I). Arithmetic mean, standard deviations, and p values (significant values are marked with an asterisk) of three repeated measurements are listed.

ROIIntegration times [seconds]0% stenosis40 % stenosis60 % stenosis80 % stenosisp-value

II
1s-2.87% (±4.85%)0.35% (±20.6%)-0.71% (±8.16%)-9.99% (±6.56%)0.71
3s18.67% (±3.2%)10.81% (±14.4%)9.68% (±4.69%)-2,73% (±4.09)0.1
5s21.12% (±5.34)11.8% (±17.03%)10.57% (±11.93%)-7.87% (±2.93%)0.06
7s19.88% (±5.61%)13.59% (±19.26%)11.17% (±13.78%)-11.07% (±2.92%)0.06
III
1s-7.25% (±3.11%)-12.64% (±20.74%)-6.06% (±19.66%)-20.83% (±6.13%)0.62
3s26.06% (±3.41%)11.73% (±10.22%)10.73% (±4.39)0.24% (±1.87%)0.0048
5s29.75% (±4.34%)7.84% (±11.3%)8.49% (±1.37%)-9.18% (±1.63%)0.0004
7s28.09% (±5.65%)5.5% (±12.78%)7.59% (±0.97%)-13.79% (±2.2%)0.0007
IV
1s-13.39% (±9.4%)-16.16% (±18.65%)-19.03% (±26.63%)-31.68% (±11.1%)0.62
3s39.38% (±5.63%)30.32% (±15.58%)12.51% (±9.96%)3.52% (±6.37%)0.0094
5s46.78% (±8.27%)25.94% (±15.81%)9.89% (±6.81%)-5.64% (±6.1%)0.0012
7s46.75% (±9.57%)23.33% (±17.65%)8.22% (±4.69%)-10.86% (±6.13%)0.0011