The Effect of Chlorhexidine and Listerine® Mouthwashes on the Tensile Strength of Selected Absorbable Sutures: An In Vitro Study

Table 1

Summary of the mean values + standard deviation (SD) of sutures tensile strength in Mega Pascal (MPa) in relation to different media and time points.

SUTUREMEDIATIME POINTS
24 hours3 days7 days10 days14 days

VICRYL 4-0DRY2151 (1257)735 (148)346 (311)755 (462)571 (708)
SALIVA1594 (805)1168 (586)1291 (454)1115 (558)565 (377)
CHX1522 (1219)1331 (681)1940 (174)1388 (470)1064 (717)
LISTERINE1345 (583)1798 (739)1580 (888)2134 (154)1984 (138)

VICRYL 5-0DRY1260 (1435)1425 (395)533 (134)938 (579)1040 (589)
SALIVA831 (448)988 (555)1025 (580)1499 (666)1011 (652)
CHX1978 (1019)2298 (1048)2045 (1031)2271 (787)2439 (177)
LISTERINE1211 (893)2057 (995)2550 (785)2656 (135)2546 (263)

MONOCRYL 4-0DRY2151 (1257)1709 (386)1291 (518)1208 (777)793 (239)
SALIVA1594 (805)1847 (831)2113 (149)1285 (574)1120 (170)
CHX1522 (1219)850 (803)756 (562)538 (312)560 (270)
LISTERINE1345 (583)1131 (741)554 (303)580 (527)352 (261)

MONOCRYL 5-0DRY1334 (943)2407 (1007)1757 (879)2255 (567)1363 (793)
SALIVA1284 (1035)2175 (1453)2121 (682)2097 (117)910 (192)
CHX2026 (1223)1546 (880)1197 (904)1143 (507)874 (548)
LISTERINE995 (406)800 (310)481 (226)437 (179)353 (178)