Research Article

A Method for Prediction of Femoral Component of Hip Prosthesis Durability due to Aseptic Loosening by Using Coffin/Manson Fatigue Model

Table 1

Calculated and recorded life of a hip prosthesis.

PatientAgeGenderBody mass

(kg)
Activity 

(hours per day)
Length 

(mm)
Normalized score of physical activity
(0-1)
Weight coefficient 

calculated
(months)

recorded (months)

140F50425.60.5374.04297.2298
247F65536.70.5454.08189.2186
353M80431.20.7035.00205.8209
455F655240.5374.04214.5205
539M783.531.10.7025.00238.1247
642F574.530.30.5714.22232.7216
741M80426.10.4143.44256.6258
842F505.531.20.5324.01205.2216
954F704300.5213.96246.1261
1048M100635.40.7125.06118.7125
1144F55536.40.5434.07205206
1250M70626.40.8415.95141.2154
1342M70529.50.5424.07195.4204
1436F60429.40.5514.11261.1278
1554F50428.80.5454.08286.4302
1637F58425.20.7025.00253253
1741F65432.40.9756.99192192
1840F45434.40.5294.00288288