A Novel Control Method for Rotary Blood Pumps as Left Ventricular Assist Device Utilizing Aortic Valve State Detection

Table 3

Accuracy of partial support and full support detection with the index according to Equation (4) at various contractility, heart rates and systemic vascular resistance in cardiovascular system; S1—Sputnik 1, S2—Sputnik 2.

Contractility, %Accuracy, %Heart rate, bpmAccuracy, %Systemic vascular resistance, %Accuracy, %
S1S2S1S2S1S2

−1510090507090−259090
−1090100608090−1590100
−5901007090100−1010090
Baseline9010080 (baseline)90100Baseline90100
59090909010015100100
1080901001009025100100
15901001101001005090100
Overall9095.7Overall88.695.7Overall94.397.1