Validity of Measurement for Trailing Limb Angle and Propulsion Force during Gait Using a Magnetic Inertial Measurement Unit
Table 1
Peak of trailing limb angle measured by magnetic inertial measurement units (IMU) and motion analysis system (mean ± SD).
| | IMU (°) | Motion analysis system (°) | ICC (95% CI) | | RMS (°) |
| Preferred | 24.1 ± 2.6 | 23.9 ± 2.6 | 0.836 (0.617–0.935) | <0.001 | 1.50 | Slow | 22.3 ± 2.7 | 22.1 ± 3.0 | 0.876 (0.699–0.952) | <0.001 | 1.42 | Fast | 26.3 ± 3.3 | 25.5 ± 3.3 | 0.833 (0.597–0.935) | <0.001 | 1.92 | Lean-preferred | 22.5 ± 3.0 | 21.8 ± 2.9 | 0.831 (0.599–0.933) | <0.001 | 1.74 | Lean-slow | 20.6 ± 3.2 | 20.0 ± 2.8 | 0.851 (0.645–0.942) | <0.001 | 1.64 |
|
|