Implementation of an Anaesthesia Quality Improvement Programme to Reduce Fibreoptic Bronchoscope Repair Incidents

Table 2

Changes in parameters related to fibreoptic bronchoscope (FOB) use, repair, cost, and unavailability before and after the implementation of the quality improvement campaign (QIC).

VariablesPre-QIC (2012)Year of QIC (2013)Follow-up
Post-QIC
Year 1 (2014)
Post-QIC
Year 2 (2015)
Post-QIC
Year 3 (2016)

Number of tracheal intubations23,45124,09324,06728,46826,327
Number of FOB55555
Number of FOI18001598166215961661
Number of repairs183033
Repair rate1%0.19%00.19%0.18%
Mean FOI per FOB360320332319332
Total repair costs (USD)18,75812,820084708137
Drop in total repair costs compared with the baseline32%100%55%57%
Cost of repairs per procedure (USD)10.428.0205.314.90
Average cost per instance of repair (USD)10424273028232712
Frequency of FOB unavailability1.4%0.4%0.3%0.3%0.3%

FOB: fibreoptic bronchoscope; FOI: fibreoptic intubation; USD: US dollars. . compared with the pre-QIC period.