Research Article

Accuracy on Scanned Images of Full Arch Models with Orthodontic Brackets by Various Intraoral Scanners in the Presence of Artificial Saliva

Table 1

Discrepancy results from superimposition of reference and scanned images (mm).

Bracket (B) Mean ± std. dev value(A) × (B)
Scanner (A)NonbracketCeramicMetalResinAverage valuesScannerBracket

MaxCS36000.60 ± 0.08cB0.61 ± 0.08cC0.96 ± 0.04aC0.86 ± 0.04bC0.758 ± 0.170C<0.001
i5000.82 ± 0.05bAB0.85 ± 0.08bBC1.15 ± 0.06aB1.08 ± 0.15aB0.975 ± 0.172B<0.001
Trios30.65 ± 0.03dB0.75 ± 0.06cBC0.97 ± 0.01bC1.05 ± 0.05aB0.854 ± 0.166C<0.001
Omnicam1.03 ± 0.37bA1.20 ± 0.29abA1.26 ± 0.03abA1.49 ± 0.13aA1.247 ± 0.286A0.002
value<0.001<0.001<0.001<0.001<0.001<0.001<0.0010.010
Average0.776 ± 0.250b0.853 ± 0.269a1.086 ± 0.132b1.119 ± 0.255a<0.001

MeanCS36000.063 ± 0.015a0.062 ± 0.014aA0.019 ± 0.008bB0.074 ± 0.008aA0.055 ± 0.024<0.001
i5000.036 ± 0.005bc0.040 ± 0.008abB0.028 ± 0.008cA0.046 ± 0.009aB0.037 ± 0.152<0.001
Trios30.056 ± 0.006b0.061 ± 0.011bA0.007 ± 0.007cC0.081 ± 0.007aA0.051 ± 0.029<0.001
Omnicam0.008 ± 0.017b0.027 ± 0.009abB0.006 ± 0.009bC0.079 ± 0.015aA0.030 ± 0.0640.031
value0.172<0.001<0.001<0.0010.579<0.001<0.0010.507
Average0.041 ± 0.061bc0.048 ± 0.019ab0.015 ± 0.018c0.070 ± 0.142a<0.001

: Max: mean maximum discrepancy values. The maximum discrepancy of the superimposed data of two images. Mean: mean discrepancy values. The average of discrepancy values throughout the 3D comparison. 2-Way ANOVA was used for statistical analysis. a >b > c > d:superscript lowercase letters indicate statistically significant differences between brackets. A > B > C > D: subscript uppercase letters indicate statistically significant differences between scanners.