Research Article

The Effect of Multiple Autoclave Sterilization on the Cyclic Fatigue of Three Heat-Treated Nickel-Titanium Rotary Files: EdgeFile X7, Vortex Blue, and TRUShape

Table 2

Tukey’s post hoc test for multiple comparisons between different groups.

TypeMean difference95% confidence interval
LowerUpper

EdgeFile (S)Vortex Blue (NS)630.0433.9826.1<0.001
Vortex Blue (S)592.6396.5788.7<0.001
TRUShape (NS)756.5560.5952.6<0.001
TRUShape (S)711.0514.9907.1<0.001
EdgeFile (NS)Vortex Blue (NS)187.1-9.0383.20.070
Vortex Blue (S)149.7-46.4345.80.234
TRUShape (NS)313.6117.5509.7<0.001
TRUShape (S)268.172.0464.10.002
Vortex Blue (S)TRUShape (NS)163.9-32.2360.00.153
TRUShape (S)118.4-77.7314.50.491
Vortex Blue (NS)TRUShape (NS)126.5-69.6322.60.415
TRUShape (S)81.0-115.1277.10.829

Significance level at .