Research Article
Preclinical Assessment of a Novel Conformable Foam-Based Left Atrial Appendage Closure Device
Table 1
Seal testing of implants in silicone tubes.
| Device | Silicone tube inner diameter (mm) | (mm) | (mm) | Seal result |
| 27 mm CLAAS | 25 | | 25 | Sealed | 27 mm CLAAS | 25 | | 25 | Sealed | 27 mm CLAAS | 25 | | 24 | Sealed | 27 mm CLAAS | 25 | | 23 | Gaps | 27 mm Watchman 2.5 | 25 | | 25 | Gaps | 27 mm Watchman 2.5 | 25 | | 25 | Gaps | 27 mm Watchman 2.5 | 25 | | 24 | Gaps | 27 mm Watchman 2.5 | 25 | | 23 | Gaps | 35 mm CLAAS | 33 | | 33 | Sealed | 35 mm CLAAS | 33 | | 32.5 | Sealed | 35 mm CLAAS | 33 | | 31.5 | Sealed | 35 mm CLAAS | 33 | | 30 | Sealed | 35 mm CLAAS | 33 | | 29.5 | Gaps |
|
|