A Mathematical Method to Adjust MLC Leaf End Position for Accurate Dose Calculation in Carbon Ion Beam Radiation Therapy Treatment Planning System

Table 2

The tungsten MLC leaf end offset correction of 400 MeV/n carbon ion beam at different fields.

Field size/2 (cm) (degree) (radian) () (radian)Projection on the -axis of ()Projection on the -axis of (light field, cm)Projection on the -axis of (MLC plan position, mm)MLC plan position shift to (mm)Light-radiation agreement (mm)Penumbra (20%-80%) (mm)

0.50.10880.00190.00190.00200.00190.50500.50.5081-0.0301-0.05041.010
10.21760.00380.00390.00390.00381.011011.0161-0.0510-0.110041.160
1.50.32640.00570.00580.00590.00581.51651.51.5241-0.0763-0.165151.310
20.43520.00760.00770.00780.00772.019722.0322-0.1248-0.197391.460
2.50.54400.00950.00960.00980.00962.52302.52.5403-0.1729-0.229781.610
30.65280.01140.01160.01180.01153.028533.0483-0.1979-0.285341.760
3.50.76160.01330.01350.01370.01343.53223.53.5564-0.2421-0.321711.910
40.87040.01520.01540.01570.01534.036844.0644-0.2767-0.367692.060
4.50.97910.01710.01740.01760.01724.54144.54.5725-0.3112-0.413672.210
51.08790.01900.01930.01960.01925.046055.0805-0.3458-0.459672.360
5.51.19670.02090.02120.02160.02115.55255.55.5886-0.3606-0.525442.510
61.30540.02280.02320.02350.02306.059866.0967-0.3689-0.597672.660
6.51.41420.02470.02510.02550.02496.56186.56.6047-0.4288-0.618312.810
71.52290.02660.02700.02740.02687.069077.1128-0.4379-0.689752.960
7.51.63160.02850.02890.02940.02887.57367.57.6208-0.4724-0.735813.110
81.74030.03040.03090.03140.03078.078288.1289-0.5068-0.781893.260
8.51.84900.03230.03280.03330.03268.58288.58.6369-0.5413-0.827983.410
91.95770.03420.03470.03530.03459.087499.1450-0.5757-0.874093.610