Research Article

In Vitro Quantitative Evaluation of Postprocessing Filter for Metal Artifact Reduction in Cone Beam Computed Tomography Images of Titanium and Zirconium Dioxide Implants

Table 1

Quantification mean (standard deviation) values of the artifact with titanium and zirconium dioxide implants.

Titanium implantZirconium dioxide
FilterMeanSDFilterMeanSD

4343
Filter 1x6.40.3Filter 1x9.20.0
Filter 2x6.30.2Filter 2x9.10.1
Filter BAR5.70.0Filter BAR7.40.1
Multi-CDT NR6.00.0Multi-CDT NR8.40.1
Nonfiltered6.70.0Nonfiltered9.50.0
4141
Filter 1x6.60.1Filter 1x9.30.2
Filter 2x6.50.2Filter 2x9.20.0
Filter BAR5.70.1Filter BAR6.70.3
Multi-CDT NR6.30.1Multi-CDT NR7.60.0
Nonfiltered6.90.2Nonfiltered9.80.1
4444
Filter 1x6.60.2Filter 1x9.20.2
Filter 2x6.50.2Filter 2x8.90.2
Filter BAR5.10.0Filter BAR7.20.1
Multi-CDT NR5.40.1Multi-CDT NR8.40.2
Nonfiltered6.50.1Nonfiltered9.10.0
4646
Filter 1x6.40.3Filter 1x8.90.0
Filter 2x6.00.2Filter 2x8.80.0
Filter BAR5.10.1Filter BAR6.60.2
Multi-CDT NR5.90.2Multi-CDT NR7.20.1
Nonfiltered6.20.2Nonfiltered9.10.2

: number of scans; SD: standard deviation.