The Effects of Additional Filtration on Image Quality and Radiation Dose in Cone Beam CT: An In Vivo Preliminary Investigation

Table 1

Filter settings and effective and relative radiation doses.

FiltersCu-filter thickness (mm)Air kerma (μGy)Relative absorbed dose (%)Interpolated effective dose (μSv)

F00 (no filter/reference)125510036†
F10.27796222.4
F20.44803813.8
F30.7328269.4
F41.0202165.8
F51.3134113.8
F61.510483.0
F72.06351.8

F0 (no filter) was used as the reference dose. Relative absorbed doses were assessed based on ion chamber dosimetric (air kerma) measurements with a repeatability error of <0.03-1%. Interpolated effective doses were calculated as linear interpolation in relation to F0. Taken from Ludlow et al. [29]. Interpolation was calculated based on relative absorbed dose measurements and a real effective dose reference value.