Research Article

Validity and Reliability of the Newly Developed Surface Electromyography Device for Measuring Muscle Activity during Voluntary Isometric Contraction

Table 2

Normalization of RMS and reliability of BTS-FREEEMG and PSL-EMG-Tr1.

VariableTrial 1Trial 2ICCDifference of means (95% CI) value

(a) Biceps brachii muscle
80% MVIC (%)BTS-FREEEMGWeek 178.92 ± 13.0084.84 ± 18.920.321−5.93 (−15.6, 3.7)0.213
Week 281.44 ± 16.9784.09 ± 19.760.185−2.65 (−14.2, 8.9)0.637
PSL-EMG-Tr1Week 176.10 ± 14.8382.08 ± 19.990.030−5.98 (−17.8, 5.9)0.306
Week 283.14 ± 15.5182.62 ± 17.930.2460.53 (−11.1, 12.2)0.925
40% MVCI (%)BTS-FREEEMGWeek 127.52 ± 10.1726.83 ± 10.890.8620.69 (−2.8, 4.2)0.683
Week 226.86 ± 9.6228.01 ± 11.220.662−1.14 (−6.1, 3.8)0.636
PSL-EMG-Tr1Week 126.07 ± 9.3924.43 ± 10.100.8311.63 (−1.9, 5.2)0.350
Week 226.07 ± 9.4928.16 ± 11.580.307−2.09 (−8.4, 4.3)0.500

(b) Rectus femoris muscle
80% MVIC (%)BTS-FREEEMGWeek 176.46 ± 18.4179.91 ± 12.590.509−3.45 (−11.9, 5.0)0.406
Week 276.70 ± 12.1978.22 ± 10.130.736−1.53 (−6.4, 3.3)0.517
PSL-EMG-Tr1Week 174.87 ± 19.2480.50 ± 11.950.547−5.63 (−13.9, 2.7)0.171
Week 277.56 ± 14.2778.81 ± 10.540.580−1.25 (−7.7, 5.2)0.688
40% MVCI (%)BTS-FREEEMGWeek 130.43 ± 7.3232.11 ± 7.780.870−1.68 (−4.0, 0.7)0.152
Week 230.07 ± 8.1831.07 ± 8.060.752−1.01 (−4.4, 2.4)0.545
PSL-EMG-Tr1Week 129.93 ± 7.9332.35 ± 8.230.765−2.42 (−5.7, 0.8)0.132
Week 230.69 ± 7.8930.84 ± 7.930.828−0.15 (−3.0, 2.7)0.912

Values are number or mean ± SD. MVIC, maximum voluntary isometric contraction; ICC, intraclass correlation coefficients; paired t-test, ; CI, confidence interval.