Analysis of the Dose Drop at the Edge of the Target Area in Heavy Ion Radiotherapy

Table 1

Dose decay data at the end edge of beam.

Parameters of TPSDose decay percentage at the end edge of beam (at a specified distance: mm)
Beam setupFractionsDose/Fx Gy (RBE)1 mm2 mm3 mm4 mm5 mm10 mm15 mm20 mm

Single direction beam20386.43%66.10%45.78%25.46%18.84%16.95%15.51%14.13%
15481.81%61.87%41.93%24.20%19.95%17.98%16.49%15.03%
10691.60%72.17%52.73%33.29%22.52%20.20%18.52%16.94%
61084.22%65.69%47.16%29.81%25.43%23.07%21.24%19.45%

Two-beam irradiation in opposite direction20392.13%78.75%65.37%51.99%45.93%41.30%38.00%35.41%
15491.66%78.57%65.47%52.38%47.02%42.40%39.10%36.49%
10693.27%80.69%68.11%55.53%49.14%44.52%41.18%38.52%
61088.25%76.49%64.74%54.45%52.07%47.60%44.31%41.65%

Horizontal and vertical (horizontal beam)20390.88%76.16%61.43%46.91%36.62%9.93%7.34%6.68%
15491.36%76.79%62.22%47.66%37.40%10.27%7.63%6.94%
10691.18%76.92%62.65%48.58%38.29%10.83%8.13%7.41%
61086.01%72.03%58.04%45.04%36.75%11.45%9.03%8.24%

Horizontal and vertical (vertical beam)20386.30%70.17%54.04%37.96%30.34%10.16%7.54%6.88%
15483.90%67.80%51.70%35.76%29.88%10.34%7.83%7.16%
10685.38%69.57%53.77%38.71%31.18%10.96%8.31%7.60%
61084.35%68.70%53.06%38.22%32.06%11.88%9.24%7.62%