Effectiveness and Safety of High-Power Radiofrequency Ablation Guided by Ablation Index for the Treatment of Atrial Fibrillation
Table 4
Procedural characteristics.
| Variable | Study group High power (40 W)
| Control group Standard power (30 W); | value |
| Procedure time, min | | | <0.001 | Right encirclement procedure time, min | | | <0.001 | Left encirclement procedure time, min | | | <0.001 | Right encirclement points, | | | 0.1 | Left encirclement points, | | | 0.407 | Total ablation time, min | | | <0.001 | Contact-force, g | | | 0.523 | Ablation time per point, s | | | <0.001 | Energy delivery per point, J | | | <0.001 | Impedance drop per point, Ω | | | <0.001 | First-pass PVI | 87 (87) | 72 (72) | <0.01 | Reconnection after 20 min | 22 (11) | 46 (23) | <0.01 | Groin hematoma | 1 | 1 | 1 | Tamponade | 0 | 0 | 1 | Periprocedural stroke | 0 | 0 | 1 | Esophageal fistula | 0 | 0 | 1 | Sinus rhythm at 12 months | 46 (92) | 42 (84) | 0.22 |
|
|
Data are presented as or (%). |