Research Article
Electromagnetic Simulation of Signal Distribution of Various RF Endoluminal Loop Geometries with Coil Orientation: Towards a Reconfigurable Design
Table 1
Tuning and matching capacitor values of each unloaded or/and loaded loop configuration.
| Loop geometry | Tuning capacitor (pF) | Matching capacitor (pF) | Ct1 | Ct2 | Cm1 | Cm2 |
| Unloaded rectangular double-loop (RDL) coils | RDL-SC | 124.88 | 124.88 | 447.31 | 447.31 | RDT-SC | 104.62 | 104.62 | 229.71 | 229.71 | RDL-OC | 157.19 | 157.19 | 424.34 | 424.34 | RDT-OC | 127.00 | 127.00 | 414.98 | — |
| Unloaded diagonal double-loop (DDL) coils | DDL-D1 | 113.75 | 113.75 | 425.00 | 425.00 | DDT-D1 | 99.36 | 99.36 | 447.78 | — | DDL-D2 | 94.95 | 94.95 | 410.00 | 410.00 | DDT-D2 | 84.75 | 84.75 | 414.18 | — |
| Loaded single and double-loop coils | DSL | — | 96.5 | 303.07 | 303.07 | RDL-OC | 157.19 | 157.19 | 424.34 | 424.34 | RDT-OC | 127.00 | 127.00 | 414.98 | — |
|
|