Research Article

Derivation and Verification of the Relationship between Ablation Index and Baseline Impedance

Table 2

The target AI, actual AI, and damage ranges obtained in different impedance under the guidance of empirical formula.

Impedance (Ω)Target AIActual AIDepths (mm)Widths (mm)

120.0 ± 1.1400402.4 ± 1.63.70 ± 0.356.30 ± 0.33
140.0 ± 0.5433433.6 ± 2.43.81 ± 0.436.07 ± 0.30
159.9 ± 0.4466463.4 ± 2.63.85 ± 0.356.11 ± 0.38
179.9 ± 0.8499497.6 ± 1.73.87 ± 0.206.18 ± 0.38
200.4 ± 0.7531533.9 ± 1.63.82 ± 0.226.09 ± 0.55