[Retracted] Effects of Medium-Term Soft Contact Lens Fitting on Dry Eye: Analyses Using Ultra-High Resolution Optical Coherence Tomography and Digital Slit-Lamp Biomicroscopy

Table 1

The repeatability of four contact lens fitting characteristics between two researchers.

VariablesPost 1 dayPost 1 weekPost 1 month
Mean differenceSD of the differencePMean differenceSD of the differencePMean differenceSD of the differenceP

Temporal lag (mm)0.0040.1190.900-0.0100.1230.770-0.0040.0740.975
Conjunctival
Coverage (mm)
-0.0030.0100.4840.0040.0140.343-0.0000.0050.836
Decentration (mm)-0.0540.1140.062-0.0290.1310.2450.0180.0790.418
Blink-induced
Movement (mm)
0.0020.0070.4650.0010.0080.863-0.0030.0250.865

SD: standard deviation.