Research Article
Microseismic Monitoring, Positioning Principle, and Sensor Layout Strategy of Rock Mass Engineering
Table 2
Technical parameters of each component.
| | Name | ESG microseismic monitoring system (MMS) |
| | Digitization | 24-bit analog-to-digital conversion | | Sensor | The sensor has a sensitivity of 30 V/g and a maximum output signal of ±5 V. 30-channel accelerometer | | Network expandable channels | Powerful integrated functions can be expanded to 256 channels | | Signal trigger mode | Threshold or STA/LTA | | Signal voltage | | | Power supply voltage | 220VAC | | Dynamic response range | >115 dB | | Data storage | Data storage can be expanded to 256 MB of 32 MB internal solid-state storage (USB, HDD can be selected), can record, and save continuous data | | Data storage format | Binary and access files to facilitate users to obtain information on up to 16 event characteristics | | Signal sampling rate | 50 Hz-10 kHz | | Signal bandwidth | DC-1/4 sampling rate | | Signal gain | 0, 6, 20, and 40 dB | | Auxiliary gain | 6-72 dB | | Energy consumption | <10 W | | Power supply | 12VDC |
|
|